清华大学学报(自然科学版)  2017, Vol. 57 Issue (7): 763-767    DOI: 10.16511/j.cnki.qhdxxb.2017.25.034
  物理与工程物理 本期目录 | 过刊浏览 | 高级检索 |
基于心冲击信号的心率检测
张先文1, 张丽岩1, 丁力超2, 魏荣荣2, 王婕1, 唐劲天1
1. 清华大学 工程物理系, 粒子技术与辐射成像教育部重点实验室, 北京 100084;
2. 东北电力大学 自动化工程学院, 吉林 132012
Heartbeat detection based on ballistocardiograms
ZHANG Xianwen1, ZHANG Liyan1, DING Lichao2, WEI Rongrong2, WANG Jie1, TANG Jintian1
1. Key Laboratory Particle & Radiation Imaging, Ministry of Education, Department of Engineering Physics, Tsinghua University, Beijing 100084, China;
2. School of Automation Engineering, Northeast Dianli University, Jilin 132012, China
全文: PDF(1170 KB)  
输出: BibTeX | EndNote (RIS)      
摘要 心冲击信号(ballistocardiogram,BCG)是由心脏搏动和大动脉血液循环引起的人体对外压力的变化,可用于无接触心脏活动检测。该文使用PVDF(polyvinylidene fluoride)压电薄膜,研制心冲击信号-心电信号(electrocardiogram,ECG)联合采集设备,并提出自适应模板匹配算法,基于心冲击信号计算心动周期,实现心率实时检测。采集24名健康成年人信号,对比ECG与BCG在心率检测方面的差异,同时评价算法的有效性。结果表明:BCG和ECG在计算心率方面无显著差异,自适应模板匹配算法查准率达到99.77%,查全率达到99.56%,准确性符合要求。
服务
把本文推荐给朋友
加入引用管理器
E-mail Alert
RSS
作者相关文章
张先文
张丽岩
丁力超
魏荣荣
王婕
唐劲天
关键词 心冲击信号(BCG)心率检测模板匹配    
Abstract:A ballistocardiogram (BCG) unobtrusively records the weight variation caused by the reaction force of blood ejection and blood flow in the aorta. A BCG system was developed to record BCG signals and detect the heart beats in the BCG signal. Data from 24 healthy subjects was collected to investigate the system with simultaneous electrocardiograms (ECG) as a reference. The results show no significant difference between the heartbeats detected by the ECG and BCG system (p=0.95) with the BCG system having a sensitivity of 99.56% and an accuracy of 99.77%.
Key wordsballistocardiogram (BCG)    heartbeat detection    templating method
收稿日期: 2016-03-07      出版日期: 2017-07-15
ZTFLH:  R331  
通讯作者: 唐劲天,研究员,E-mail:tangjt@tsinghua.edu.cn     E-mail: tangjt@tsinghua.edu.cn
引用本文:   
张先文, 张丽岩, 丁力超, 魏荣荣, 王婕, 唐劲天. 基于心冲击信号的心率检测[J]. 清华大学学报(自然科学版), 2017, 57(7): 763-767.
ZHANG Xianwen, ZHANG Liyan, DING Lichao, WEI Rongrong, WANG Jie, TANG Jintian. Heartbeat detection based on ballistocardiograms. Journal of Tsinghua University(Science and Technology), 2017, 57(7): 763-767.
链接本文:  
http://jst.tsinghuajournals.com/CN/10.16511/j.cnki.qhdxxb.2017.25.034  或          http://jst.tsinghuajournals.com/CN/Y2017/V57/I7/763
  图1 BCGGECG 联合采集系统框图
  图2 3个不同健康成年人BCG 模板
  图3 模板匹配过程图示
  图4 系统采集到某样本一段BCG 与ECG 信号
  表1 BCG 与ECG 计算心动周期检验结果
  图5 BCG 与ECG 计算得到心动周期对比图
  图6 自适应模板匹配算法计算结果
  表2 24名健康成年人自适应模板匹配算法计算结
[1] Starr I, Schroeder H A. Ballistocardiogram. Ⅱ. Normal standards, abnormalities commonly found in diseases of the heart and circulation, and their significance [J]. Journal of Clinical Investigation, 1940, 19(3): 437-450.
[2] Gordon J. Certain molar movements of the human body produced by the circulation of the blood [J]. Journal of Anatomy and Physiology, 1877, 11(3): 533-536.
[3] Starr I, Rawson A, Schroeder H, et al. Studies on the estimation of cardiac ouptut in man, and of abnormalities in cardiac function, from the heart's recoil and the blood's impacts: The ballistocardiogram [J]. American Journal of Physiology—Legacy Content, 1939, 127(1): 1-28.
[4] Inan O, Etemadi M, Wiard R, et al. Robust ballistocardiogram acquisition for home monitoring [J]. Physiological Measurement, 2009, 30(2): 169-185.
[5] Gómez Clapers J, Serra Rocamora A, Casanella R, et al. Uncertainty factors in time-interval measurements in ballistocardiography [C]//Proceedings of the 19th IMEKO TC-4 Symposium 17th IWADC Workshop Adverbial Instrum-entation and Sensors Interoperability. Barcelona, Spain: IMEKO, 2013: 395-399.
[6] Junnila S, Akhbardeh A, Värri A. An electromechanical film sensor based wireless ballistocardiographic chair: Implementation and performance [J]. Journal of Signal Processing Systems, 2009, 57(3): 305-320.
[7] Jansen B H, Larson B H, Shankar K. Monitoring of the ballistocardiogram with the static charge sensitive bed [J]. IEEE Transactions on Biomedical Engineering, 1991, 38(8): 748-751.
[8] Chuo Y, Marzencki M, Hung B, et al. Mechanically flexible wireless multisensor platform for human physical activity and vitals monitoring [J]. IEEE Transactions on Biomedical Circuits and Systems, 2010, 4(5): 281-294.
[9] Choi B H, Chung G S, Lee J S, et al. Slow-wave sleep estimation on a load-cell-installed bed: A non-constrained method [J]. Physiological Measurement, 2009, 30(11): 1163-1170.
[10] Kortelainen J M, Virkkala J. FFT averaging of multichannel BCG signals from bed mattress sensor to improve estimation of heart beat interval [C]//Engineering in Medicine and Biology Society, 2007, 29th Annual International Conference of the IEEE. Lyon, France: IEEE, 2007: 6685-6688.
[11] Etemadi M, Inan O T, Giovangrandi L, et al. Rapid assessment of cardiac contractility on a home bathroom scale [J].IEEE Transactions on Information Technology in Biomedicine, 2011, 15(6): 864-869.
[12] Giovangrandi L, Inan O T, Banerjee D, et al. Preliminary results from BCG and ECG measurements in the heart failure clinic [C]//Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE. San Diego, CA, USA: IEEE, 2012: 3780-3783.
No related articles found!
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
版权所有 © 《清华大学学报(自然科学版)》编辑部
本系统由北京玛格泰克科技发展有限公司设计开发 技术支持:support@magtech.com.cn