Please wait a minute...
 首页  期刊介绍 期刊订阅 联系我们
 
最新录用  |  预出版  |  当期目录  |  过刊浏览  |  阅读排行  |  下载排行  |  引用排行  |  百年期刊
Journal of Tsinghua University(Science and Technology)    2017, Vol. 57 Issue (7) : 763-767     DOI: 10.16511/j.cnki.qhdxxb.2017.25.034
PHYSICS AND ENGINEERING PHYSICS |
Heartbeat detection based on ballistocardiograms
ZHANG Xianwen1, ZHANG Liyan1, DING Lichao2, WEI Rongrong2, WANG Jie1, TANG Jintian1
1. Key Laboratory Particle & Radiation Imaging, Ministry of Education, Department of Engineering Physics, Tsinghua University, Beijing 100084, China;
2. School of Automation Engineering, Northeast Dianli University, Jilin 132012, China
Download: PDF(1170 KB)  
Export: BibTeX | EndNote | Reference Manager | ProCite | RefWorks    
Abstract  A ballistocardiogram (BCG) unobtrusively records the weight variation caused by the reaction force of blood ejection and blood flow in the aorta. A BCG system was developed to record BCG signals and detect the heart beats in the BCG signal. Data from 24 healthy subjects was collected to investigate the system with simultaneous electrocardiograms (ECG) as a reference. The results show no significant difference between the heartbeats detected by the ECG and BCG system (p=0.95) with the BCG system having a sensitivity of 99.56% and an accuracy of 99.77%.
Keywords ballistocardiogram (BCG)      heartbeat detection      templating method     
ZTFLH:  R331  
Issue Date: 15 July 2017
Service
E-mail this article
E-mail Alert
RSS
Articles by authors
ZHANG Xianwen
ZHANG Liyan
DING Lichao
WEI Rongrong
WANG Jie
TANG Jintian
Cite this article:   
ZHANG Xianwen,ZHANG Liyan,DING Lichao, et al. Heartbeat detection based on ballistocardiograms[J]. Journal of Tsinghua University(Science and Technology), 2017, 57(7): 763-767.
URL:  
http://jst.tsinghuajournals.com/EN/10.16511/j.cnki.qhdxxb.2017.25.034     OR     http://jst.tsinghuajournals.com/EN/Y2017/V57/I7/763
  
  
  
  
  
  
  
  
[1] Starr I, Schroeder H A. Ballistocardiogram. Ⅱ. Normal standards, abnormalities commonly found in diseases of the heart and circulation, and their significance [J]. Journal of Clinical Investigation, 1940, 19(3): 437-450.
[2] Gordon J. Certain molar movements of the human body produced by the circulation of the blood [J]. Journal of Anatomy and Physiology, 1877, 11(3): 533-536.
[3] Starr I, Rawson A, Schroeder H, et al. Studies on the estimation of cardiac ouptut in man, and of abnormalities in cardiac function, from the heart's recoil and the blood's impacts: The ballistocardiogram [J]. American Journal of Physiology—Legacy Content, 1939, 127(1): 1-28.
url: http://dx.doi.org/can Journal of Physiology
[4] Inan O, Etemadi M, Wiard R, et al. Robust ballistocardiogram acquisition for home monitoring [J]. Physiological Measurement, 2009, 30(2): 169-185.
[5] Gómez Clapers J, Serra Rocamora A, Casanella R, et al. Uncertainty factors in time-interval measurements in ballistocardiography [C]//Proceedings of the 19th IMEKO TC-4 Symposium 17th IWADC Workshop Adverbial Instrum-entation and Sensors Interoperability. Barcelona, Spain: IMEKO, 2013: 395-399.
[6] Junnila S, Akhbardeh A, Värri A. An electromechanical film sensor based wireless ballistocardiographic chair: Implementation and performance [J]. Journal of Signal Processing Systems, 2009, 57(3): 305-320.
[7] Jansen B H, Larson B H, Shankar K. Monitoring of the ballistocardiogram with the static charge sensitive bed [J]. IEEE Transactions on Biomedical Engineering, 1991, 38(8): 748-751.
[8] Chuo Y, Marzencki M, Hung B, et al. Mechanically flexible wireless multisensor platform for human physical activity and vitals monitoring [J]. IEEE Transactions on Biomedical Circuits and Systems, 2010, 4(5): 281-294.
[9] Choi B H, Chung G S, Lee J S, et al. Slow-wave sleep estimation on a load-cell-installed bed: A non-constrained method [J]. Physiological Measurement, 2009, 30(11): 1163-1170.
[10] Kortelainen J M, Virkkala J. FFT averaging of multichannel BCG signals from bed mattress sensor to improve estimation of heart beat interval [C]//Engineering in Medicine and Biology Society, 2007, 29th Annual International Conference of the IEEE. Lyon, France: IEEE, 2007: 6685-6688.
[11] Etemadi M, Inan O T, Giovangrandi L, et al. Rapid assessment of cardiac contractility on a home bathroom scale [J].IEEE Transactions on Information Technology in Biomedicine, 2011, 15(6): 864-869.
[12] Giovangrandi L, Inan O T, Banerjee D, et al. Preliminary results from BCG and ECG measurements in the heart failure clinic [C]//Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE. San Diego, CA, USA: IEEE, 2012: 3780-3783.
[1] Xinrong CAO, Lei LIU, Dongyang CAI, Peng GUO, Jintian TANG. Statistical analyses of ballistocardiogram features for cardiac disease diagnosis[J]. Journal of Tsinghua University(Science and Technology), 2014, 54(5): 633-637.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
Copyright © Journal of Tsinghua University(Science and Technology), All Rights Reserved.
Powered by Beijing Magtech Co. Ltd