Performance evaluation of a PET detector witha sparse SiPM array and gap reflectors |
WEI Qingyang1,2, WANG Fenhua1, XU Tianpeng3, JIA Chao3, JIANG Nianming3, GU Yu1, MA Tianyu2, LIU Yaqiang2 |
1. Beijing Engineering Research Center of Industrial Spectrum Imaging, School of Automation and Electrical Engineering, University of Science and Technology Beijing, Beijing 100083, China; 2. Key Laboratory of Particle & Radiation Imaging(Tsinghua University), Ministry of Education, Beijing 100084, China; 3. Beijing Novel Medical Equipment Ltd., Beijing 102206, China |
|
|
Abstract Positron emission tomography (PET) usually uses scintillation detectors consisting of crystal arrays and photoelectric sensor arrays. In recent years, silicon photomultipliers (SiPMs) have been more widely used in PET detectors. Although a closely arranged SiPM array can give high performance, a sparse array gives a more cost-effective system. This paper describes a sparse 8×8 SiPM array using the MicroFB-30035-SMT chip from SensL Inc. The array size is 33.7 mm×33.7 mm while the SiPM chip is 3.16 mm×3.16 mm; thus, the gap ratio is 44%. The sparse SiPM array and a dual-layer offset LYSO array were used in a high performance depth PET detector. The gaps between the SiPMs are dead zones which reduce the optical photon collection. Thus, this work studies the effects of adding enhanced spectral reflector films into the gaps. Flood maps were acquired with and without reflectors at room temperature with the crystal response analyzed to find the photopeaks, energy resolution and root-mean square (RMS) of the crystal response. The results show that the reflectors in the gaps effectively enhance the photon collection (25.5% increase) and optimize the energy resolution of the detector (from 13.48% to 12.80%). The quality of the flood map is also improved, i.e. the intrinsic spatial resolution of the PET detector.
|
Keywords
positron emission tomography (PET)
sparse SiPM array
reflector
performance evaluation
|
Issue Date: 17 October 2018
|
|
|
[1] MUEHLLEHNER G, KARP J S. Positron emission tomography[J]. Physics in Medicine & Biology, 2006, 51(13):R117-R137. [2] WEI Q Y, WANG S, DAI T T, et al. SiPM based PET detector modules with air-gapped pixelated LYSO[C]//Proceedings of the 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference. Seattle, WA, USA:IEEE, 2014:1-3. [3] ZHANG H, ZHOU R, YANG C. A PET detector module with monolithic crystal, single end readout, SiPM array and high depth-of-interaction resolution[J]. Journal of Instrumentation, 2016, 11(8):P08020. [4] ZHANG X M, WANG X H, REN N, et al. Performance of a SiPM based semi-monolithic scintillator PET detector[J]. Physics in Medicine & Biology, 2017, 62(19):7889-7904. [5] GOERTZEN A L, STORTZ G, THIESSEN J D, et al. First results from a high-resolution small animal SiPM PET insert for PET/MR imaging at 7T[J]. IEEE Transactions on Nuclear Science, 2016, 63(5):2424-2433. [6] LEVIN C S, MARAMRAJU S H, KHALIGHI M M, et al. Design features and mutual compatibility studies of the time-of-flight PET capable GE SIGNA PET/MR system[J]. IEEE Transactions on Medical Imaging, 2016, 35(8):1907-1914. [7] LEE J H, LEE S J, AN S J, et al. Development of a SiPM-based PET detector using a digital positioning algorithm[J]. Journal of the Korean Physical Society, 2016, 68(9):1049-1054. [8] LI B C, WANG Y, XI D M, et al. Feasibility study on silicon photomultiplier with epitaxial quenching resistors as the readout for PET detectors[J]. IEEE Transactions on Nuclear Science, 2016, 63(1):17-21. [9] 魏清阳, 戴甜甜, 谷宇, 等. 基于蛇形光路的PET探测器作用深度提取方法[J]. 光子学报, 2017, 46(3):0304002. WEI Q Y, DAI T T, GU Y, et al. Extraction of the depth-of-interaction for PET detectors based on a serpentine-light-path design[J]. Acta Photonica Sinica, 2017, 46(3):0304002. (in Chinese) [10] WANG Y J, ZHANG Z M, LI D W, et al. Design and performance evaluation of a compact, large-area PET detector module based on silicon photomultipliers[J]. Nuclear Instruments and Methods in Physics Research Section A:Accelerators, Spectrometers, Detectors and Associated Equipment, 2012, 670:49-54. [11] RONCALI E, CHERRY S R. Application of silicon photomultipliers to positron emission tomography[J]. Annals of Biomedical Engineering, 2011, 39(4):1358-1377. [12] ZHANG Y X, YAN H, BAGHAEI H, et al. A novel depth-of-interaction block detector for positron emission tomography using a dichotomous orthogonal symmetry decoding concept[J]. Physics in Medicine & Biology, 2016, 61(4):1608-1633. [13] WEI Q Y, XU T P, DAI T T, et al. Development of a compact DOI-TOF detector module for high-performance PET systems[J]. Nuclear Science and Techniques, 2017, 28(4):43. [14] SensL. J Series (High Performance/TSV)-Datasheet[EB/OL].[2017-12-12]. http://sensl.com/documentation. [15] ZHU X Z, DENG Z, CHEN Y, et al. Development of a 64-channel readout ASIC for an 8×8 SSPM array for PET and TOF-PET applications[J]. IEEE Transactions on Nuclear Science, 2016, 63(3):1327-1334. [16] WEI Q Y, DAI T T, MA T Y, et al. Crystal identification in dual-layer-offset DOI-PET detectors using stratified peak tracking based on SVD and mean-shift algorithm[J]. IEEE Transactions on Nuclear Science, 2016, 63(5):2502-2508. [17] WEI Q Y, MA T Y, XU T P, et al. Evaluation of signal energy calculation methods for a light-sharing SiPM-based PET detector[J]. Nuclear Instruments and Methods in Physics Research Section A:Accelerators, Spectrometers, Detectors and Associated Equipment, 2017, 848:81-86. |
|
Viewed |
|
|
|
Full text
|
|
|
|
|
Abstract
|
|
|
|
|
Cited |
|
|
|
|
|
Shared |
|
|
|
|
|
Discussed |
|
|
|
|