Neutron multiplicity counter design and calibration
Ligao CHEN1,2,Xiaobo LIU2,Jian GONG2(),Kan WANG1,Chuanjiang DONG2,Ru RONG2,Yi LU2
1. Department of Engineering Physics, Tsinghua University, Beijing 100084, China
2. Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China
Download: PDF(1305 KB)   HTML
Export: BibTeX | EndNote | Reference Manager | ProCite | RefWorks     Supporting Info
Guide   
Abstract  

A neutron multiplicity counter is designed and calibrated with the integrated electronics data acquisition and processing software described here. The predelay time length search equations are developed by balancing the correlated counts and the background counts of Am-Li neutron sources. A deadtime coefficient fitting method minimizes the root mean square. Nine parameters are acquired and two plutonium samples are used to verify the counter performance and parameter calibration algorithm. Test results show that the design and calibration for neutron multiplicity measurements are valid.

Keywords neutron multiplicity      counter building      calibration      deadtime fitting     
ZTFLH:     
Fund: 
Issue Date: 15 February 2014
Service
E-mail this article
E-mail Alert
RSS
Articles by authors
Ligao CHEN
Xiaobo LIU
Jian GONG
Kan WANG
Chuanjiang DONG
Ru RONG
Yi LU
Cite this article:   
Ligao CHEN,Xiaobo LIU,Jian GONG, et al. Neutron multiplicity counter design and calibration[J]. Journal of Tsinghua University(Science and Technology), 2014, 54(2): 159-163.
URL:  
http://jst.tsinghuajournals.com/EN/     OR     http://jst.tsinghuajournals.com/EN/Y2014/V54/I2/159
  
  
  
  
252Cf
编号
Sm/s S0/s Dm/s D0/s Sm/Dm S0/D0
B≠0 B=0 B≠0 B=0 B≠0 B=0
1# 63.343 63.341 63.345 5.959 5.958 5.959 10.630 10.631 10.629
2# 81.415 81.412 81.418 7.760 7.759 7.761 10.492 10.493 10.490
3# 142.360 142.350 142.370 13.697 13.693 13.701 10.394 10.396 10.392
4# 709.750 709.510 709.980 70.795 70.701 70.886 10.025 10.035 10.016
5# 2 135.201 2 133.402 2 137.305 202.862 202.171 203.643 10.525 10.552 10.495
6# 19 387.123 19 550.235 19 557.316 1 814.721 1 876.522 1 879.432 10.683 10.418 10.406
  
252Cf编号 Tm/s T0/s Sm/Tm S0/T0
1# 0.332 0.349 191 181
2# 0.433 0.456 188 179
1#+2# 0.752 0.792 189 180
3# 3.860 4.160 184 171
4# 9.290 10.800 230 197
5# 10.200 110.011 1 910 179
  
252Cf编号 计数率/cps fd ft τ/μs
1# 63.3 0.520 67 0.303 15 58.69
2# 81.4 0.527 56 0.307 76 58.38
3# 142.4 0.532 61 0.305 87 58.35
4# 709.6 0.552 57 0.322 44 58.27
5# 2 135.2 0.527 31 0.279 01 58.11
6# 19 387.3 0.531 85 0.308 18 53.79
  
Pu编号 测量时间/h m/g
(真实值)
m/g
(测量值)
氧化物的质量分数
(真实值)
氧化物的质量分数
(测量值)
RSD/%
1# 3 496.2 502.8 0 -0.02 0.88
2# 6 2 561.0 2 302.3 0 -0.02 1.32
  
[1] Macarthur D W, Whiteson R, Langner D, et al. Attribute Measurement System with Information Barrier for the Fissile Material Transparency Technology Demonstration System Overview, LA-UR-99-5611 [R]. Los Alamos, NM: Los Alamos National Laboratory, 1999.
[2] Chambers D, David M. UK-Norway Initiative: Research into information barriers to allow warhead attribute verification without release of sensitive or proliferative information [C]//Proceedings of the 51st Annual Meeting of the Institute of Nuclear Materials Management. Baltimore: INMM Press, 2010: 201-211.
[3] 贾向军, 王效忠, 许小明. 中子多重性测量数据获取软件的开发 [C]//中国原子能科学研究院年报. 北京: 原子能出版, 2001: 116. JIA Xiangjun, WANG Xiaozhong, XU Xiaoming. Development of neutron multiplicity dada acquisition software [C]//Annals of the China Institute of Atomic Energy. Beijing: Atomic Energy Press, 2001: 116. (in Chinese)
[4] 王中杰, 张全虎, 刘杰, 等. 基于MCNP模拟的中子多重性计数器3He的设计[J]. 核电子学与探测技术, 2010, 30(12): 626-628. WANG Zhongjie, ZHANG Quanhu, LIU Jie, et al.Designing of the 3He detector in neutron multiplicity counter based on MCNP simulation[J]. Nuclear Electronics & Detection Technology, 2010, 30(12): 626-628. (in Chinese)
[5] 黎素芬, 张全虎, 弟宇鸣, 等. 有源中子多重性测量计算机模拟 [C]//第16届全国核电子学与核探测技术学术年会论文集. 北京: 中国电子学会出版社, 2012: 105-113. LI Sufen, ZHANG Quanhu, DI Yuming, et al.Active neutron multiplicity stochastic simulation research [C]//The 16th National Nuclear Electronics and Nuclear Detection Technology Academic Meeting. Beijing: Chinese Electronics Press, 2012: 105-113. (in Chinese)
[6] Rinard P M, Krick M S, Kelley T A, et al. Measurements on an Inventory of Mixed Fissile Materials in Shipping Containers, LA-13356-MS [R]. Los Alamos, NM: Los Alamos National Laboratory, 1997.
[7] Swansen J E, Collinsworth P R, Krick M S, et al. Multiplicity Sorter for Shift-Register Coincidence Electronics, LA-9221-MS [R]. Los Alamos, NM: Los Alamos National Laboratory, 1982.
[8] Dytlewski N. Dead-time corrections for multiplicity counters[J]. Nuclear Instrument and Method, 1991, A305: 492-494.
[9] Dytlewski N, Krick M S, Ensslin N. Measurement variances in thermal neutron coincidence counting[J]. Nuclear Instrument and Method in Physics, 1993, A327: 469-479.
[10] Krick M S, Swansen J E. Neutron multiplicity and multiplication measurements[J]. Nuclear Instrument and Method, 1984, A219: 384-393.
[11] Peerani P, Weber A, Swinhoe M T, et al.ESARDA multiplicity benchmark exercise[J]. Esarda Bulletin, 2009, 34: 3-32.
[12] 胡大璞, 郑福裕. 核反应堆物理实验方法 [M]. 北京: 原子能出版社, 1988. HU Dapu, ZHENG Fuyu. Method of Reactor Physics Experiments [M]. Beijing: Atomic Energy Press, 1988. (in Chinese)
[1] FENG Yuxing, ZHENG Jun, LIN Jinsong. Calibration method of laser profiler based on constant focusing optical path[J]. Journal of Tsinghua University(Science and Technology), 2024, 64(4): 738-748.
[2] WU Shan, WU Yuchen, HOU Benwei, HAN Hongquan. Water demand dynamic estimation in water distribution network hydraulic models based on Kalman filter[J]. Journal of Tsinghua University(Science and Technology), 2024, 64(2): 271-281.
[3] LI Zhongwei, LI Xiaofei, Tang Zuozhou, XU Wenting, SONG Qiang. Optimization of thermogravimetric method for measuring very low saturation vapor pressure[J]. Journal of Tsinghua University(Science and Technology), 2023, 63(11): 1833-1843.
[4] LI Zhengqing, HOU Senhao, WEI Jinhao, TANG Xiaoqiang. Vision-based auto-calibration method for planar cable-driven parallel robot for warehouse and logistics tasks[J]. Journal of Tsinghua University(Science and Technology), 2022, 62(9): 1508-1515.
[5] ZHU Chuanhui, ZHU Zhiming, KE Zhijie, ZHANG Tianyi. Integrated calibration of internal visual sensor parameters based on combined laser structured lights[J]. Journal of Tsinghua University(Science and Technology), 2022, 62(9): 1516-1523.
[6] HU Congjun, YU Guang, WANG Liping. Calibration of a 5-axis hybrid machine based on a 3P(4R)S spindle head[J]. Journal of Tsinghua University(Science and Technology), 2019, 59(12): 1029-1038.
[7] ZHENG Jun, LI Wenqing. Calibration of 3-D measurement system based on a double position sensitive detectors[J]. Journal of Tsinghua University(Science and Technology), 2018, 58(4): 411-416.
[8] LIU Yu, ZHANG Xiujun, ZHOU Shidong. Reciprocity calibration of a massive MIMO based on a local calibration[J]. Journal of Tsinghua University(Science and Technology), 2017, 57(9): 939-944,951.
[9] FENG Chuqing, SHEN Qi, KANG Kejun, XING Yuxiang. Energy calibration of photon counting detectors using a single monochromatic source[J]. Journal of Tsinghua University(Science and Technology), 2016, 56(6): 672-676.
[10] ZHANG Xinxi, ZHANG Rong, GUO Meifeng, CHENG Gaofeng, NIU Shulai. Online three-axis magnetometer calibration for a pedestrian navigation system using a foot-mounted inertial navigation system[J]. Journal of Tsinghua University(Science and Technology), 2016, 56(2): 211-217.
[11] LIU Yuzhe, WU Jun, WANG Liping, WANG Jinsong. Measurement trajectory evaluation and error compensation for kinematic calibration of a 5-axis hybrid machine tool[J]. Journal of Tsinghua University(Science and Technology), 2016, 56(10): 1047-1054.
[12] Mengjiao WANG,Hui DING,Guangzhi WANG. Misaligned cone beam computed tomography reconstruction based on a camera model[J]. Journal of Tsinghua University(Science and Technology), 2015, 55(1): 122-127.
[13] Xiao LUO, Yuan YAO, Jinhuan ZHANG. Unified calibration method for millimeter-wave radar and camera[J]. Journal of Tsinghua University(Science and Technology), 2014, 54(3): 289-293.
[14] Jinhuan ZHANG, Weiguo LIU, Jingtao LI, Fuquan ZHAO. Pedestrian headform impact test FE modeling and analysis of parameter sensitivity[J]. Journal of Tsinghua University(Science and Technology), 2014, 54(3): 294-298.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
Copyright © Journal of Tsinghua University(Science and Technology), All Rights Reserved.
Powered by Beijing Magtech Co. Ltd