Journal of Tsinghua University(Science and Technology)    2014, Vol. 54 Issue (2) : 202-206     DOI:
Orginal Article |
Improved resolution for SPR detection systems with angular modulation
Fuyin DOU,Peng WANG,Xinglong YU()
State Key Laboratory of Precision Measurement Technology and Instruments, Department of Precision Instruments, Tsinghua University, Beijing 100084, China
Download: PDF(2415 KB)   HTML
Export: BibTeX | EndNote | Reference Manager | ProCite | RefWorks     Supporting Info
Guide   
Abstract  

Improved system resolutions for surface plasmon resonance (SPR) will enable applications in more fields. This paper shows theoretically that the resolution of SPR with angular modulation is not only improved with increased sensitivity, but also with a narrower half-peak width. The resolutions with Au, Ag and Au-Ag chips show that the resolution of the Au-Ag chip is higher than that of the Au chip. Thus, the system resolution can be improved by using the Au-Ag chip. Measurements of the resolution of Au and Au-Ag chips using NaCl solutions with different mass fractions show that the Au and Au-Ag chip resolutions are 5.2×10-6 and 3.9×10-6 RIU (refractive index unit). Thus, the Au-Ag chip resolution is 1.4 times higher than the Au chip, similar to the theoretical result being 1.6 times higher.

Keywords surface plasmon resonance (SPR)      system resolution      sensitivity      Au-Ag chip     
ZTFLH:     
Fund: 
Issue Date: 15 February 2014
Service
E-mail this article
E-mail Alert
RSS
Articles by authors
Fuyin DOU
Peng WANG
Xinglong YU
Cite this article:   
Fuyin DOU,Peng WANG,Xinglong YU. Improved resolution for SPR detection systems with angular modulation[J]. Journal of Tsinghua University(Science and Technology), 2014, 54(2): 202-206.
URL:  
http://jst.tsinghuajournals.com/EN/     OR     http://jst.tsinghuajournals.com/EN/Y2014/V54/I2/202
  
  
  
  
dAg/nm,
dAu/nm
Sθ/((°)·RIU-1) Wθ/(°) σRI(归一化)
48,0 135 3.5 0.34
37,5 148 5.0 0.45
27,15 152 7.0 0.63
16,25 160 10.0 0.83
0,41 161 12.0 1.00
  
  
  
  
[1] Myszka D G, Rebecca L R. Implementing surface plasmon resonance biosensors in drug discovery[J]. Pharmaceutical Science & Technology Today, 2000, 3(9): 310-316.
[2] Workman P, Aboagye E, Balkwill F, et al.Guidelines for the welfare and use of animals in cancer research[J]. British Journal of Cancer, 2010, 102: 1555-1577.
url: http://dx.doi.org/10.1038/sj.bjc.6605642
[3] Mulchan N M, Rodriguez M, Kevin O, et al.Application of a high-resolution SPR technique for monitoring real-time metal/dielectric interactions[J]. Sensors and Actuators B, 2003, 88: 132-137.
url: http://dx.doi.org/10.1016/S0925-4005(02)00316-7
[4] 定翔, 余兴龙, 耿俊清. 表面等离子体共振生物传感技术及仪器化的过去、 现在与未来[J]. 现代科学仪器, 2012, 3: 124-134. DING Xiang, YU Xinglong, GENG Junqing. Development history of surface plasmon resonance biosensor technology and instrumentation[J]. Modern Scientific Instruments, 2012, 3: 124-134. (in Chinese)
url: http://www.cnki.com.cn/Article/CJFDTotal-XDYQ201203041.htm
[5] Zynio S A, Samoylov A V, Surotseva E R, et al.Bimetallic layers increase sensitivity of affinity sensor based on surface plasmon resonance[J]. Sensors, 2002, 2: 62-70.
[6] Biow H O, YUAN Xiaocong, Tjin S C, et al.Optimised film thickness for maximum evanescent fild enhancement of a bimetallic film surface plasmon resonance biosensor[J]. Sensors and Actuators B, 2006, 114: 1028-1034.
url: http://dx.doi.org/10.1016/j.snb.2005.07.064
[7] Johansen K, Stalbert R, Lundstrom I, et al.Surface plasmon resonance: Instrumental resolution using photo diode array[J]. Measurement Science and Technology, 2000, 11: 1630-1638.
url: http://dx.doi.org/10.1088/0957-0233/11/11/313
[8] Tao N J, Boussad S, Huang W L, et al.High resolution surface plasmon resonance spectroscopy[J]. Review of Scientific Instrument, 1999, 70(12): 4656-4600.
url: http://dx.doi.org/10.1063/1.1150128
[9] Ekgasit S, Thammacharoen C, YU Fang, et al.Evanescent field in surface plasmon resonance and surface plasmon field-enhanced fluorescence spectroscopies[J]. Analytical Chemistry, 2004, 76(8): 2210-2219.
url: http://dx.doi.org/10.1021/ac035326f
[10] Piliarik M, Homola J. Surface plasmon resonance (SPR) sensors: Approaching their limits?[J]. Optics Express, 2009, 17: 16505-16517.
url: http://dx.doi.org/10.1364/OE.17.016505
[11] Homola J, Koudela I, Yee S S. Surface plasmon resonance sensors based on diffraction gratings and prism couplers: Sensitivity comparison[J]. Sensors and Actuators B, 1999, 54: 16-24.
url: http://dx.doi.org/10.1016/S0925-4005(98)00322-0
[12] Bruijn H E, Kooyman R P H, Greve J. Choice of metal and wavelength for surface-plasmon resonance sensors: Some considerations[J]. Applied Optics, 1992, 31(4): 440-442.
url: http://dx.doi.org/10.1364/AO.31.000440
[13] Ekgasit S, YU Fang, Knoll W. Fluorescence intensity in surface-plasmon field-enhanced fluorescence spectroscopy[J]. Sensors and Actuators B, 2005, 104: 294-301.
url: http://dx.doi.org/10.1016/j.snb.2004.05.021
[14] 耿俊清, 余兴龙, 章恩耀, 等. 基于相位解算的SPR传感方法及其应用[J]. 仪器仪表学报, 2012, 33(10): 15-21. GENG Junqing, YU Xinglong, ZHANG Enyao, et al.SPR biosensing method based on phase shift algorithm and its application[J]. Chinese Journal of Scientific Instrument, 2012, 33(10): 15-21. (in Chinese)
url: http://www.cnki.com.cn/Article/CJFDTotal-YQXB201210002.htm
[1] WANG Zhexin, LIU Hui, CHENG Li, GAO Lilei, LV Zhenlei, JIANG Nianming, HE Zuoxiang, LIU Yaqiang. Design of dedicated collimator for whole-body bone scanning on single photon emission computed tomography based on Monte Carlo simulation[J]. Journal of Tsinghua University(Science and Technology), 2023, 63(5): 811-817.
[2] LIU Peng, QIAO Xinzhou. Stability sensitivity of a completely restrained 3-DOF cable-driven parallel robot with four long-span cables[J]. Journal of Tsinghua University(Science and Technology), 2022, 62(9): 1548-1558.
[3] MA Linwei, YUAN Yuan, LI Zheng. Mapping the characteristics and sensitivities of China's low-carbon energy supply in 2050[J]. Journal of Tsinghua University(Science and Technology), 2022, 62(4): 802-809.
[4] ZHOU Changcong, CHANG Qi, ZHOU Chunping, ZHAO Haodong, SHI Zhuangke. Fault tree analysis of an aircraft flap system based on a non-probability model[J]. Journal of Tsinghua University(Science and Technology), 2021, 61(6): 636-642.
[5] WANG Wandi, ZHANG Hui. Influence of smoke exhaust start-up time on smoke control[J]. Journal of Tsinghua University(Science and Technology), 2016, 56(2): 223-228.
[6] YU Hao, GONG Hui, WANG Yongqiang, WANG Xuewu, LI Jianmin, LI Yuanjing, KANG Kejun. Readout electronics for position sensitive boron coated straw detector[J]. Journal of Tsinghua University(Science and Technology), 2015, 55(6): 691-694.
[7] LIU Hongyu, XU Yuejin, JIANG Peiyan. Capital cost sensitivity of real estate corporate investment[J]. Journal of Tsinghua University(Science and Technology), 2015, 55(6): 640-646.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
Copyright © Journal of Tsinghua University(Science and Technology), All Rights Reserved.
Powered by Beijing Magtech Co. Ltd