Average vector method for determining isocenter of rotated radiotherapy equipment
ZHANG Xiang, FAN Wengang, SHI Shaohua, YE Peiqing, ZHANG Hui
Beijing Key Laboratory of Precision/Ultra-Precision Manufacturing Equipments and Control, State Key Laboratory of Tribology, Institute of Manufacture Engineering, Department of Mechanical Engineering, Tsinghua University, Beijing 100084, China
Download: PDF(1511 KB)  
Export: BibTeX | EndNote | Reference Manager | ProCite | RefWorks    
Abstract  The isocentric accuracy of rotated radiotherapy equipment significantly influences the treatment accuracy. The average vector method is used to find the optimal isocenter considering the mechanical deformation of the rotated radiotherapy equipment. The maximum distance between the isocenter which is treated as the optimization variable and the ray vectors is considered as a constraint. An objective function is then designed to calculate the average distance between the isocenter that meets the constraint and the ray vectors. Finally, the location of the optimal isocenter and the isocentric envelop are obtained. A test with this method gives the radius of the isocentric envelop as 0.311 mm which is equal to that given by the vector-end-effector method, and about 15.5% less than that given by the point-end-effector method. The results also show that the average distance between the optimal isocenter and the ray vectors is 0.206 mm, about 11.2% less than that by the vector-end-effector method and 28% less than that by the point-end-effector method.
Keywords radiotherapy equipment      isocenter      ray vector      average vector method     
ZTFLH:  TH692.9  
Issue Date: 15 July 2015
Service
E-mail this article
E-mail Alert
RSS
Articles by authors
ZHANG Xiang
FAN Wengang
SHI Shaohua
YE Peiqing
ZHANG Hui
Cite this article:   
ZHANG Xiang,FAN Wengang,SHI Shaohua, et al. Average vector method for determining isocenter of rotated radiotherapy equipment[J]. Journal of Tsinghua University(Science and Technology), 2015, 55(7): 744-749.
URL:  
http://jst.tsinghuajournals.com/EN/     OR     http://jst.tsinghuajournals.com/EN/Y2015/V55/I7/744
   
   
   
   
   
   
   
   
   
   
[1] 李少林, 吴永忠. 肿瘤放射治疗学[M]. 北京: 科学出版社, 2013: 1-3.LI Shaolin, WU Yongzhong. Tumor Radiation Therapy [M]. Beijing: Science Press, 2013: 1-3. (in Chinese)
[2] 胡逸民. 肿瘤放射物理[M]. 北京: 原子能出版社, 1999.HU Yimin. Radiation Oncology Physics [M]. Beijing: Atomic Energy Press, 1999.(in Chinese)
[3] 顾本广. 医用加速器[M]. 北京: 科学出版社, 2003: 341-356.GU Benguang. Medical Accelerator [M]. Beijing: Science Press, 2003: 341-356. (in Chinese)
[4] 黄嘉华. 放疗设备等中心精度分析与对策[J]. 上海生物医学工程, 2003, 24(1): 10-13.HUA Jiahua. Isocenter rotation resolution analysis and technique of radiotherapy equipment [J].Journal of Shanghai Biomedical Engineering Society, 2003, 24(1): 10-13. (in Chinese)
[5] 胡杰, 陶建民, 孙光荣, 等. 直线加速器等中心的质量保证和质量控制[J]. 中国医疗器械杂志, 2007, 31(3): 213-215.HU Jie, TAO Jianmin, SUN Guangrong, et al. Quality assurance and quality control of accelerator isocenter [J]. Chinese Medical Journal, 2007, 31(3): 213-215. (in Chinese)
[6] 刘丕福, 钟毓斌. 等中心放射治疗技术[J]. 中日友好医院学报, 1988, 2(2): 111-114.LIU Pifu, ZHONG Yubin. Isocentric radiation therapy technology [J]. Journal of China-Japan Friendship Hospital, 1988, 2(2): 111-114. (in Chinese)
[7] DU Weiliang, GAO Song, WANG Xiaochun, et al. Quantifying the gantry sag on linear accelerators and introducing an MLC-based compensation strategy [J]. Med Phys, 2012, 39(4): 2156-2162.
[8] DU Weiliang, GAO Song. Measuring the wobble of radiation field centers during gantry rotation and collimator movement on a linear accelerator [J]. Med Phys, 2011, 38(8): 4575-4578.
[9] Gonzalez A, Castro I, Martinez J A. A procedure to determine the radiation isocenter size in a linear accelerator [J]. Med Phys, 2004, 31(6): 1489-1493.
[10] Tsai J S, Curran B H, Sternick E S, et al. The measurement of linear accelerator isocenter motion using a three-micometer device and an adjustable pointer [J]. Int J Radiat Oncol Biol Phys, 1996, 34(1): 189-195.
[11] Skworcow P, Mills J A, Haas O C, et al. A new approach to quantify the mechanical and radiation isocentres of radiotherapy treatment machine gantries [J]. Phys Med Biol, 2007, 52(23): 7109-7124.
[12] Riis H L, Zimmermann S J, Hjelm-Hansen M. Gantry and isocenter displacements of a linear accelerator caused by an add-on micromutileaf collimator [J]. Med Phys, 2013, 40(3): 031707-1-031707-9.
[13] Rosca F, Lorenz F, Hacker F L, et al. An MLC-based linac QA procedure for the characterization of radiation isocenter and room laser's position [J]. Med Phys, 2006, 33(6): 1780-1787.
No related articles found!
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
Copyright © Journal of Tsinghua University(Science and Technology), All Rights Reserved.
Powered by Beijing Magtech Co. Ltd