心冲击信号(ballistocardiogram,BCG)是由心脏搏动和大动脉血液循环引起的人体对外压力的变化,可用于无接触心脏活动检测。该文使用PVDF(polyvinylidene fluoride)压电薄膜,研制心冲击信号-心电信号(electrocardiogram,ECG)联合采集设备,并提出自适应模板匹配算法,基于心冲击信号计算心动周期,实现心率实时检测。采集24名健康成年人信号,对比ECG与BCG在心率检测方面的差异,同时评价算法的有效性。结果表明:BCG和ECG在计算心率方面无显著差异,自适应模板匹配算法查准率达到99.77%,查全率达到99.56%,准确性符合要求。
A ballistocardiogram (BCG) unobtrusively records the weight variation caused by the reaction force of blood ejection and blood flow in the aorta. A BCG system was developed to record BCG signals and detect the heart beats in the BCG signal. Data from 24 healthy subjects was collected to investigate the system with simultaneous electrocardiograms (ECG) as a reference. The results show no significant difference between the heartbeats detected by the ECG and BCG system (p=0.95) with the BCG system having a sensitivity of 99.56% and an accuracy of 99.77%.
[1] Starr I, Schroeder H A. Ballistocardiogram. Ⅱ. Normal standards, abnormalities commonly found in diseases of the heart and circulation, and their significance [J]. Journal of Clinical Investigation, 1940, 19(3): 437-450. [2] Gordon J. Certain molar movements of the human body produced by the circulation of the blood [J]. Journal of Anatomy and Physiology, 1877, 11(3): 533-536. [3] Starr I, Rawson A, Schroeder H, et al. Studies on the estimation of cardiac ouptut in man, and of abnormalities in cardiac function, from the heart's recoil and the blood's impacts: The ballistocardiogram [J]. American Journal of Physiology—Legacy Content, 1939, 127(1): 1-28. [4] Inan O, Etemadi M, Wiard R, et al. Robust ballistocardiogram acquisition for home monitoring [J]. Physiological Measurement, 2009, 30(2): 169-185. [5] Gómez Clapers J, Serra Rocamora A, Casanella R, et al. Uncertainty factors in time-interval measurements in ballistocardiography [C]//Proceedings of the 19th IMEKO TC-4 Symposium 17th IWADC Workshop Adverbial Instrum-entation and Sensors Interoperability. Barcelona, Spain: IMEKO, 2013: 395-399. [6] Junnila S, Akhbardeh A, Värri A. An electromechanical film sensor based wireless ballistocardiographic chair: Implementation and performance [J]. Journal of Signal Processing Systems, 2009, 57(3): 305-320. [7] Jansen B H, Larson B H, Shankar K. Monitoring of the ballistocardiogram with the static charge sensitive bed [J]. IEEE Transactions on Biomedical Engineering, 1991, 38(8): 748-751. [8] Chuo Y, Marzencki M, Hung B, et al. Mechanically flexible wireless multisensor platform for human physical activity and vitals monitoring [J]. IEEE Transactions on Biomedical Circuits and Systems, 2010, 4(5): 281-294. [9] Choi B H, Chung G S, Lee J S, et al. Slow-wave sleep estimation on a load-cell-installed bed: A non-constrained method [J]. Physiological Measurement, 2009, 30(11): 1163-1170. [10] Kortelainen J M, Virkkala J. FFT averaging of multichannel BCG signals from bed mattress sensor to improve estimation of heart beat interval [C]//Engineering in Medicine and Biology Society, 2007, 29th Annual International Conference of the IEEE. Lyon, France: IEEE, 2007: 6685-6688. [11] Etemadi M, Inan O T, Giovangrandi L, et al. Rapid assessment of cardiac contractility on a home bathroom scale [J].IEEE Transactions on Information Technology in Biomedicine, 2011, 15(6): 864-869. [12] Giovangrandi L, Inan O T, Banerjee D, et al. Preliminary results from BCG and ECG measurements in the heart failure clinic [C]//Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE. San Diego, CA, USA: IEEE, 2012: 3780-3783.